Agilent -N9038A MXE EMI receiver
Meet CISPR 16-1-1 2010 EMI receiver requirements
Absolute amplitude accuracy of ±0.75 dB at 1 GHz
The preamplifier can be used to generate -163 dBm display average noise level (DANL)
When the 1 GHz center frequency is 10 kHz offset, the phase noise is -106 dBc/Hz
Function
Frequency 20 Hz ~ 8.4 GHz and 26.5 GHz
CISPR bandwidth 200 Hz, 9 kHz, 120 kHz, and 1 MHz; MIL-STD and bandwidth increments in 10% steps
Various diagnostic tools, including EMC measurement applications
Perfectly switch between EMI receiver and spectrum analyzer modes
performance
Meet CISPR 16-1-1 2010 EMI receiver requirements
Absolute amplitude accuracy of ±0.75 dB at 1 GHz
The preamplifier can be used to generate -163 dBm display average noise level (DANL)
When the 1 GHz center frequency is 10 kHz offset, the phase noise is -106 dBc/Hz
Measurement applications and software
EMC measurement application (as standard) can provide real-time detector, signal list, scan table, limit line and correction factor
SCPI command language compatibility (as standard) can simulate R&S FSP/FSU/FSE spectrum analyzers
Support phase noise measurement application software and other measurement application software to be provided
Use the cursor function, zoom span, area span, and spectrum to enhance your analysis capabilities
Automatic test and communication interface
Comply with LXI C standard, and support SCPI and IVI-COM
USB 2.0, 1000Base-T LAN, GPIB
Universal X series user interface / open Windows? XP operating system